Anesthetic Gas Measurement; Spectroscopy) - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Technical data
Anesthetic gas measurement (infrared spectroscopy)
Measuring ranges of the anesthetic gases:
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
Accuracy of anesthetic gas measurement:
Resolution of anesthetic gas measurement:
N
O
2
xMAC
All gases
Anesthetic gas detection
Cross sensitivity
3 ‰ alcohol
1 ‰ acetone or methane
100 ppm NO or CO
Water steam saturated at 37 °C
412
Measuring range: 0 to 8.5 Vol.%
Measuring range: 0 to 10 Vol.%
Measuring range: 0 to 8.5 Vol.%
Measuring range: 0 to 10 Vol.%
Measuring range: 0 to 20 Vol.%
±(0.2 Vol.% + 15 % of the measured value) in ac-
cordance with ISO 21647
Drift of measuring accuracy is compensated by cy-
clic zeroing.
0.1 Vol.%
Measuring range: 0 to 100 Vol.%
Accuracy: ±(2.0 Vol.% + 8 % of the measured
value) in accordance with ISO 21647
Resolution: 1 Vol.%
Measuring range: 0 to 9.9 xMAC
Resolution: 0.01 xMAC
Display: based on the expiratory measured values
Alarms: based on the inspiratory measured values
Automatic
Primary gas: from 0.3 Vol.%
Secondary gas: from 0.4 Vol.%
Distortion <0.2 Vol.% for anesthetic gas, CO
N
O measurement.
2
Distortion <1 Vol.% for O
Instructions for use Zeus Infinity Empowered SW 2.n
, and
2
measurement.
2