Dräger Zeus Infinity Empowered Instructions For Use Manual page 307

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Classifi-
Pri-
Alarm text
cation
ority
Caution
250
Simulated patient
data
Warning
000
Sinus bradycardia
Warning
000
Sinus bradycardia
Warning
120
SmartCare can-
celed
Note
100
Sound system er-
ror
Caution
125
SpO
Note
100
SpO
Warning
250
SpO
Note
250
SpO
Note
150
SpO
nected?
Note
100
SpO
invalid
Instructions for use Zeus Infinity Empowered SW 2.n
Cause
Patient data is simulated by
software.
8 or more consecutive nor-
mal beats, with an average
rate < sinus bradycardia rate
setting. In neonatal mode:
heart rate lower BRDY-Set-
ting
8 or more consecutive nor-
mal beats, with an average
rate < sinus bradycardia rate
setting. In neonatal mode:
heart rate lower BRDY-Set-
ting
Patient values are not within
the etCO
technical problem occured.
The sound system failed.
There will be no audible
alarms.
high
SpO
2
upper alarm limit.
HW failure
SpO
2
hardware failure.
low
SpO
2
lower alarm limit.
pulse search
The monitor has not been
2
able to compute a valid
measurement within the last
30 seconds because of un-
stable measurement condi-
tions.
sensor con-
Too much or insufficient light
2
is reaching the sensor or
measurement artefact Weak
signal due to poor perfusion
or low body temperature.
sensor
The monitor does not recog-
2
nize the connected sensor
as valid.
and VT ranges. A
2
is above the current
2
measurement system
2
is below the current
2
Fault – Cause – Remedy
Remedy
Use simulated patient data
for demonstration purposes
only.
Check patient and take ap-
propriate action. Check Ar-
rhythmia settings. Neona-
tal: Check BRADY settings.
Check patient and take ap-
propriate action. Check Ar-
rhythmia settings. Neona-
tal: Check BRADY settings.
Check patient condition.
Remedy technical alarms.
Check ventilation settings.
Only use system in case of
emergency. Observe visual
messages at all times.
Observe the patient and
take appropriate action.
Call DrägerService.
Observe the patient and
take appropriate action.
Ensure the SpO
sensor is
2
properly attached to the pa-
tient and all cables are
properly connected. Call
DrägerService.
Check sensor attachment
Check patient's condition.
Check sensor and cables.
Relocate sensor to another
extremity.
Check for defective or
unapproved sensor.
Replace the sensor. Call
DrägerService.
307