Classifi-
Pri-
Alarm text
cation
ority
Warning
150
Battery very low
Caution
100
BIS high
Caution
100
BIS low
Caution
050
BIS out of range
Caution
050
BIS out of range
Note
210
BIS signal quality
poor
Note
050
BIS/NMT alarms
off
Instructions for use Zeus Infinity Empowered SW 2.n
Cause
The capacity of the uninter-
ruptible power supply is ex-
hausted (<10 %).
BIS value above higher
alarm limit. Patients hypnotic
state not appropriate.
BIS value below lower alarm
limit.
BIS parameter value out of
range.
BIS parameter value out of
range.
SQI (Signal Quality Index)
value goes below 50 %.
BIS and NMT alarms sus-
pended
Fault – Cause – Remedy
Remedy
Restore power supply. Pre-
pare for man. ventilation.
Minimize power consump-
tion. Check battery capacity
display. If capacity is <10 %
the device may be switched
off!
Check patient. Treat, if nec-
essary. Increase agent
dose if necessary. Change
alarm limit as appropriate.
Check patient. Treat, if nec-
essary. Decrease agent
dose if necessary. Change
alarm limit as appropriate.
Check patient and treat, if
necessary.
Check patient and treat, if
necessary.
Check sensor. This may oc-
cur as a result of artefact
caused e.g. by eyeblinks or
motion. Isolate patient from
auxiliary equipment. Isolate
patient from extraneous
movement or vibration.
Check patient. Treat, if nec-
essary.
Watch measured values to
monitor patients status.