Cleaning, disinfection, and sterilization
3 Slide the new disposable CO
holder (D).
4 Press the CO
absorber into the device (D) until
2
it engages. See also: page 67.

Attaching the breathing hoses

WARNING
Check the breathing hoses for damage before
each use. Do not use damaged breathing
hoses.
For operation in a closed system, hose water traps
can be used in order to collect the condensate (at-
tach at the lowest point).
C
1 Attach the breathing hoses (A) to the inspiration
and expiration nozzles of the breathing system
and connect to the Y-piece.
2 Attach the HME filter (B) or filter to the nozzle of
the Y-piece.
3 Establish the Luer-Lock connection (C) of the
sample line between the HME filter of the Y-
piece and the CO
2
4 Connect the breathing bag (D) with the connec-
tion nozzle to the breathing hose. Attach the
shorter breathing hose to the angled nozzle and
attach the breathing bag to the bag holder so
that it hangs down.
380
absorber on the
2
A
A
D
C
water trap.
For perfect operation in a closed system, the use of
Dräger breathing bags (see list of accessories) is
recommended.
WARNING
Do not use any antistatic or conductive
breathing hoses!
Risk of burns when using HF surgery equip-
ment.
NOTE
Zeus IE is manufactured without natural rubber
latex.
To minimize the risk of exposure to latex, use
breathing bags and breathing hoses manufac-
tured without natural rubber latex.

Assembling the anesthetic gas receiving

system (AGS)

B
1 Install particle filter, tighten the union nut (A).
2 Screw on the flow tube (B).
3 Tighten the union nut (C).
4 Reattach the buffer-volume container (D).
Instructions for use Zeus Infinity Empowered SW 2.n
C
B
A
D