Variant 2 and variant 3

D
O
Air N
O
O
Air N
O
2
2
2
2
C
E F G
All the controls are on the screen.
A The screen is touch-sensitive (touch screen)
and can also be operated with gloves. Touch
lightly to call up functions or dialog windows.
B Rotary knob for selecting and confirming set-
tings
C On/Off key to switch the device on/off
D LED for status display of the power supply for
grid operation:
– Green = mains supply sufficient
– Off = no mains supply present
E LED for status display of the power supply for
battery operation:
– Green = battery charged
– Yellow = battery charge less than 50 %
– Off = battery charge less than 10 %
F LEDs for status display of the gas supply from
the central gas supply (O
– Green = pressure sufficient
– Off = pressure not sufficient
Instructions for use Zeus Infinity Empowered SW 2.n
A
B
, Air, N
O)
2
2
G LEDs for status display of the gas supply from
the backup gas cylinders (O
– Green = pressure sufficient
I
– Green, flashing = pressure slightly below
level
– Off = pressure not sufficient
H Alarm pause function
Press the
alert of all active alarms for 2 minutes. For addi-
tional information, see page 177.
I
The alarm bar on the front and back indicates
alarm conditions:
H
– Red = high-priority alarm condition
– Yellow = medium-priority alarm condition
Operating concept
, Air, N
2
2
button to suppress the audible
O)
43