Cleaning, disinfection, and sterilization
Replacing the water trap
1 Remove hose (A) and pull off water trap (B).
2 Label the new water trap on the marked field
with the current date.
3 Push the water trap (B) into the holder again un-
til it engages.
Connection for internal sample line to
the Protect water trap
Connection for sample line to the
CO
water trap
2
CAUTION
Do not put the device into operation without a wa-
ter trap. Water and bacteria can get into the meas-
uring system. Contamination has a negative ef-
fect on gas analysis, which can fail completely.
Risk of device contamination.
CAUTION
Do not spray the O-rings of the water trap holder
with silicon spray. Silicone may get into the meas-
uring cell and distort gas analysis.
362
Removing the CO
Reusable absorber
B
1 Turn the CO
absorber (A) counterclockwise
2
and remove downwards.
2 Pull the IBF absorber filter and absorber insert
(B) out of the CO
absorber, leaving the internal
2
and external sealing rings on the absorber in-
sert.
3 Empty the soda lime, see also page 66. Ob-
serve the instructions for use of the absorber.
Dispose of the soda lime and IBF absorber filter
according to the relevant instructions for use.
4 Prepare the CO
absorber for cleaning and dis-
2
infection in the cleaning and disinfection ma-
chine.