Classifi-
Pri-
Alarm text
cation
ority
Note
080
Pump 1 infusion
limit
Caution
150
Pump 1 infusion
limit
Caution
200
Pump 1 operation
failed
Caution
060
Pump 1 pressure
drop
Caution
060
Pump 1 pressure
rise
Note
125
Pump 1 removed
Note
125
Pump 1 removed
Note
100
Pump 1 unlocked
Caution
040
Pump 2 battery
empty
Note
040
Pump 2 battery low
Caution
200
Pump 2 failure
Note
120
Pump 2 incompati-
ble
Instructions for use Zeus Infinity Empowered SW 2.n
Cause
Infusion limit is nearly
reached for pump 1. If limit is
reached infusion will be
stopped!
Upper alarm limit for infusion
volume 1 exceeded. Infusion
stopped.
Remote control, setting or
start/stop of pump 1 failed.
Pressure drop in syringe 1
detected. Infusion line is
leaking.
Pressure rise in syringe 1
detected. Infusion line
kinked or blocked.
Pump 1 removed during run-
ning infusion.
Pump 1 removed in stopped
mode.
Pump 1 is unlocked.
Infusion 2 stopped. Battery
of pump 2 discharged.
Battery nearly empty.
Infusion 2 stopped. Pump
detected internal failure.
Pump 2 hardware or soft-
ware version not compatible.
Remote control not possible.
Fault – Cause – Remedy
Remedy
Prepare next syringe if nec-
essary. Increase alarm limit
if needed.
Check syringe 1. Check
alarm limits and increase
limit if necessary.
Check last setting of pump
1. Repeat setting or start of
pump, if necessary.
Check infusion line and
connections at patient and
syringe side. Be aware of
leakages. Deactivate the
DPS- function at the pump.
Check infusion line 1 and
syringe 1. Deactivate the
DPS- function at the pump,
if appropriate.
Control infusion at the
pump. To restart remote
control reconnect pump.
To restart remote control re-
connect pump.
Lock the pump.
Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Exchange pump 2. Call
DrägerService.
Control infusion at pump.
Replace pump for remote
control.
299