Alarm
T1/T2 (°C)
TBlood (°C)
TBlood (°C)
TOF-Count
TOF-Count
BIS
BIS

Arrhythmia detection

Alarm
Setting range
ASY
Alarm
On/Off
Rate
Count
VF
Alarm
On/Off
Rate
Count
V.tach
Alarm
On/Off
Rate
100 to 200
Count
5 to 15
V.run
Alarm
On/Off
Rate
100 to 200
Count
3 to (VTcount –1)
AIVR
Alarm
On/Off
Rate
100 to 200
(<VTcount –1)
Count
count >3
182
Setting range
–5 to 49.9
25.1 to 43
25 to 42.9
1 to 4; Off
0 to 3; Off
15 to 100; Off
0 to 95; Off
Resolu-
tion/incre-
ment
10
1
10
1
10
Default value of factory profile
Adult 2 Adult 1
Default value of factory profile
Adult 2
Adult 1
Ped 2
On
On
On
On
On
On
≥120
≥120
≥10
≥10
On
On
≥120
≥120
3 to 9
3 to 9
3 to 9
On
On
119
119
3
3
Instructions for use Zeus Infinity Empowered SW 2.n
Ped 2
Ped 1
34
39
34
4
0
100
10
Ped 1
Neo 1
On
On
On
On
On
On
≥120
≥120
≥10
≥10
On
On
≥120
≥120
3 to 9
On
On
119
119
3
3
Neo 1
Off
Off
Off
Off
Off

Hide quick links: