Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Caution
125
TOF count high
Caution
250
TOF count low
Note
100
Too many pumps
Note
100
Trident pod discon-
nected
Note
105
Trident pod hard-
ware fail
Note
150
Two BISx pods de-
tected
Note
100
Two NMT pods de-
tected
Note
150
Two SpO
detected
Note
000
Unable to learn
Note
020
V electrode fail
Note
020
V+ electrode fail
314
Cause
TOF count is higher than the
the set alarm limit.
TOF count is lower than the
set alarm limit.
Maximum number of pumps
exceeded. System only sup-
ports up to 4 pumps.
Trident pod not connected to
system.
Trident pod hardware failure.
An unsupported pod is con-
nected.
A second BISx Pod is con-
nected.
A second NMT pod is con-
nected.
sensors
A second SpO
2
connected.
No dominant complex for
QRS processing found after
100 beats and until the end
of learning phase. Learning
will be continued.
Lead-off condition for indi-
cated lead detected: -Broken
cable – Loose lead wire
-Faulty lead wire – Wrong
lead -Dried out gel -Bad con-
tact.
Lead-off condition for indi-
cated lead detected: -Broken
cable - Loose lead wire
-Faulty lead wire - Wrong
lead -Dried out gel -Bad con-
tact.
Remedy
Check patient, treat if nec-
essary. Change alarm limit
as appropriate.
Check patient, treat if nec-
essary. Change alarm limit
as appropriate.
Only 4 pumps are displayed
and can be controlled.
Check cables and connec-
tions. Replace cables, if
necessary.
Check Trident pod
connection. Disconnect
unsupported pod and
reconnect Trident pod. Call
DrägerService.
Remove the BISx pod that
was connected last.
Remove the NMT pod that
was connected last.
sensor is
Remove the SpO
2
connected last.
Check electrode applica-
tion. Check leads. Choose
another lead for QRS
processing.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Instructions for use Zeus Infinity Empowered SW 2.n