Operation

Influence of weight and age on the device settings:

Patient data
Influenced parameter
Weight
Tidal volume: VT
Respiratory rate: RR
∆Psupp cancelation
Internal respiratory phase
detection
Alarm limit setting for
Exp. time too short
Internal circle flow in
fresh-gas mode
Age
xMAC

Integrated SmartPilot View (optional)

SmartPilot View* is only available in the adult
(Adult) patient category.
B
Touch the SmartPilot button (A).
*
only in conjunction with Variant 3 – C700
114
Algorithm
Radford / User / VT from profile
Radford / User / RR from profile
>20 kg: 4.0 s
≤20 kg: 1.5 s
>20 kg: Criteria for adults
≤20 kg: Criteria for pediatric pa-
tients
>20 kg: 6.0 L/min
≤20 kg: 2.0 L/min
>20 kg: 6 L/min x T
/(T
i
≤20 kg: 1 L/min x T
/(T
i
Age-dependent after Mapleson
Integrated SmartPilot View is optionally displayed
or hidden in the left-hand area (B) of the screen.
The view can be saved in a profile (Profiles).
1:35
More detailed information can be found in the
SmartPilot View instructions for use (9054084,
1st edition).
A
Instructions for use Zeus Infinity Empowered SW 2.n
Remark
Setting is performed
only in standby mode.
Setting is performed
only in standby mode.
+ T
)
i
e
+ T
)
i
e