Operation
Special events during ventilation
Water trap is full
Remove water trap.
Empty and replace water trap, see page 361.
Soda lime is used up
Remove absorber container in fresh-gas or
standby mode.
Replace soda lime, see page 66.
DIVA metering module is empty
Refill the DIVA metering module, see page 75.
Switch over to full DIVA metering module or
Change DIVA metering module, see page 74.
Suction system is full
Change suction bottle.
Insert empty backup bottle (see relevant in-
structions for use).
134
Changing the patient
Patient data can only be configured in the Startup
dialog (see page 109) dialog from the standby
screen, from a ventilation mode, and from Vent.
Standby using the New patient button.
NOTE
Starting a new case clears the data of the preced-
ing case!
Changing the patient from ventilation mode
System standby?
B
1 Touch the Start/Standby button (A).
All settings previously modified in operation are
preserved.
The System standby? dialog window will be
opened.
2 Confirm Stop ventilation? button (B) with ro-
tary knob.
Or continue ventilation:
Close dialog window.
Pressing the rotary knob automatically switches the
device to the system standby screen.
3 Call up the Startup dialog and configure,
see page 109.
Instructions for use Zeus Infinity Empowered SW 2.n