USB devices
The General settings > USB devices page opens.
Additional components can be installed. The
Zeus IE supports mass storage and input devices.
CAUTION
Unsupported USB devices
USB devices that are not supported or have too
high a current consumption may impair correct
device operation.
Use only supported USB devices with a maximum
current consumption of 500 mA.
CAUTION
USB devices with their own mains power supply
USB devices with their own mains power supply
(e.g., USB printers) may endanger patients and
users.
Connect only USB devices that do not have their
own mains power supply to Zeus IE.
CAUTION
Computer viruses
Mass storage devices can contain computer vi-
ruses which can impair the operation of Zeus IE.
– Check mass storage devices for computer vi-
ruses before using them with Zeus IE.
– Do not use any mass storage device contain-
ing computer viruses with Zeus IE.
Basic settings
A
Instructions for use Zeus Infinity Empowered SW 2.n
Starting the installation:
1 Touch the Install new device button (A).
2 Connect the USB device to be installed.
Observe the message on the screen.
After successful installation, the The following de-
vices have been detected message appears on
the screen.
Configuration