Dräger Zeus Infinity Empowered Instructions For Use Manual page 311

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Classifi-
Pri-
Alarm text
cation
ority
Caution
050
Syringe 2 nearly
empty
Caution
125
Syringe 2 unlocked
Warning
130
Syringe 3 empty
Caution
050
Syringe 3 nearly
empty
Caution
125
Syringe 3 unlocked
Warning
130
Syringe 4 empty
Caution
050
Syringe 4 nearly
empty
Instructions for use Zeus Infinity Empowered SW 2.n
Cause
Syringe 2 nearly empty.
5 min until end of infusion or
10 % volume left. (depend-
ing on infusion rate and type
of syringe)
Infusion 2 can not start while
clasp is unlocked.
Infusion 3 stopped.
Syringe 3 empty.
Syringe 3 nearly empty.
5 min until end of infusion or
10 % volume left. (depend-
ing on infusion rate and type
of syringe)
Infusion 3 can not start while
clasp is unlocked.
Infusion 4 stopped.
Syringe 4 empty.
Syringe 4 nearly empty.
5 min until end of infusion or
10 % volume left. (depend-
ing on infusion rate and type
of syringe)
Fault – Cause – Remedy
Remedy
Control syringe. Prepare
filled syringe, if necessary.
Lock clasp at the pump 2.
Control right position of the
syringe. and start infusion
or bolus.
Control syringe and patient.
Replace syringe, if neces-
sary. To drain syringe com-
pletely, confirm the green
button at the pump. (Only
possible, if 'Empty syringe
mode' is configured).
Control syringe. Prepare
filled syringe, if necessary.
Lock clasp at the pump 3.
Control right position of the
syringe. and start infusion
or bolus.
Control syringe and patient.
Replace syringe, if neces-
sary. To drain syringe com-
pletely, confirm the green
button at the pump. (Only
possible, if 'Empty syringe
mode' is configured).
Control syringe. Prepare
filled syringe, if necessary.
311