Dräger Zeus Infinity Empowered Instructions For Use Manual page 410

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Technical data
Measuring system (Cont.)
Minute volume MV (spontaneous or mandatory)
Respiratory rate (RR or RRspon)
Patient compliance Cdyn
Patient resistance R
O
flow tube
2
Response time of the whole system
(according to ISO 80601-2-55)
410
Measuring range: 0 to 40 L/min
Resolution: 0.1 L/min
Accuracy: ±10 % (under calibration conditions)
T
value on the display: approx. 45 s
90
Measuring range: 0 to 150 1/min
Resolution: 1 1/min
Accuracy: ±1 1/min
T
value on the display: approx. 45 s
90
Measuring range: 0 to 300 mL/hPa
Resolution: 1 mL/hPa
Qualitative value only!
Average: 10 breaths
Measuring range: 0 to 200 hPa/L/s
Resolution: 1 hPa/L/s
Qualitative value only!
Average: 10 breaths
Setting range: 0 to 16 L/min
Accuracy: ±10 % of the set value
4 +2 seconds