Section 2: Introduction, Features, And Specifications; Introduction; Warmer Module; Cart/Bassinet (Resuscitaire Birthing Room Warmer (Wbr82) And Cart Only) - Dräger Resuscitaire WBR82 Instructions For Use Manual

Radiant warmer product
Hide thumbs Also See for Resuscitaire WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04

Introduction

There are 3 versions of the Resuscitaire Radiant Warmer product: the Resuscitaire Birthing Room
Warmer (WBR82), the Resuscitaire Radiant Warmer (RW82), and the Resuscitaire Wall Mounted
Radiant Warmer (WMRW82).

Warmer Module

A controller provides Pre-Warm Mode, Manual Mode, or Baby Mode (servo skin temperature control) to
control the warmer. An examination light provides additional illumination of the mattress area. A warmer
head pivot allows the warmer to be pivoted to either side for x-ray procedures. Additionally, when the
warmer is pivoted, it continues to provide heat.

Cart/Bassinet (Resuscitaire Birthing Room Warmer (WBR82) and Cart only)

The bassinet is designed to provide maximum function and utility to aid in the care of the newborn. It is
detachable so you can transport the infant to the NICU, the general nursery, or another area of the
hospital. The side and front panels fold down to permit access to the infant. The mattress tilts at a 0 °, 5 °,
or 10 ° angle. Openings are provided on each side of the bassinet for the insertion of the accessory x-ray
tray. A writing surface is available; when the writing surface is closed, it is not visible. Each cart has a
pass-through drawer and a locking front access drawer. The cart locks in place when used with a warmer.

Bassinet (Resuscitaire Radiant Warmer (RW82) only)

The bassinet is designed to provide maximum function and utility to aid in the care of the newborn. Fold
the side and front panels down to gain access to the infant. The mattress tilt is continually adjustable at
±10 °. Openings are provided on each side of the bassinet for the insertion of the accessory x-ray tray. A
push-through drawer, 2 instrument trays, anti-static castors, a steering castor, and a cylinder mounting
bracket are supplied standard on every unit.

Controller

At power-up, the microprocessor within the controller performs a series of diagnostic tests to confirm the
proper operation of the system. During this time (approximately 6 to 10 secs.), all display and indicators
light, except for Power Fail, and a tone sounds. At the end, the alarm sounds a high-pitched tone, a low-
pitched tone, and then 3 beeps.
When powered-up, the system initializes in Pre-Warm Mode. The controller starts the heater at 100%
power, and maintains that setting for 3 min. Then it reduces to 60% for 12 min, and then reduces the
heater power to 30%. During the low heat setting, the unit continues to provide heat with no low
temperature alarms activated.
Introduction, Features,
and Specifications
Section 2
2 - 1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Resuscitaire rw82Resuscitaire wmrw82

Table of Contents