Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Note
080
Pump 4 infusion
limit
Caution
150
Pump 4 infusion
limit
Caution
200
Pump 4 operation
failed
Caution
060
Pump 4 pressure
drop
Caution
060
Pump 4 pressure
rise
Note
125
Pump 4 removed
Note
125
Pump 4 removed
Note
100
Pump 4 unlocked
Caution
000
PVC high
Note
050
RA check zero
Note
020
RA electrode fail
302
Cause
Infusion limit is nearly
reached for pump 4. If limit is
reached infusion will be
stopped!
Upper alarm limit for infusion
volume 4 exceeded. Infusion
stopped.
Remote control, setting or
start/stop of pump 4 failed.
Pressure drop in syringe 4
detected. Infusion line is
leaking.
Pressure rise in syringe 4
detected. Infusion line
kinked or blocked.
Pump 4 removed during run-
ning infusion.
Pump 4 removed in stopped
mode.
Pump 4 is unlocked.
Premature ventricular con-
tractions: PVCs per minute
have exceeded the upper
alarm limit.
Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Lead-off condition for indi-
cated lead detected: -Broken
cable - Loose lead wire -
Faulty lead wire - Wrong lead
-Dried out gel -Bad contact.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Prepare next syringe if nec-
essary. Increase alarm limit
if needed.
Check syringe 4. Check
alarm limits and increase
limit if necessary.
Check last setting of pump
4. Repeat setting or start of
pump, if necessary.
Check infusion line and
connections at patient and
syringe side. Be aware of
leakages. Deactivate the
DPS- function at the pump.
Check infusion line 4 and
syringe 4. Deactivate the
DPS- function at the pump,
if appropriate.
Control infusion at the
pump. To restart remote
control reconnect pump.
To restart remote control re-
connect pump.
Lock the pump.
Check patient and take ap-
propriate action.
Zero the transducer.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.