Dräger Zeus Infinity Empowered Instructions For Use Manual page 284

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Caution
000
Insp. oxygen high
Warning
180
Insp. oxygen low
Note
175
Insp. pressure sen-
sor fail
Caution
160
Insp. resistance
high
Caution
255
Insp. sevoflurane
high
Caution
255
Insp. sevoflurane
high
Note
255
Insp. sevoflurane
high
Caution
060
Insp. sevoflurane
low
Caution
254
Insp. vol. agent
high
284
Cause
Upper limit of insp. Oxygen-
concentration exceeded for
2 breath. cycles.
Inspiratory Oxygen concen-
tration fell below lower limit
for 2 breath. cycles.
Failure of insp. pressure
sensor. Measurement of
breathing pressure may be
inaccurate. Breathing press.
is measured by exp. sensor.
Inspiratory resistance is in-
creased compared to self-
test – Stenosis or high resist-
ance in insp. hose, tube or fil-
ter.
Insp. vol. agent higher than
alarm limit for 30 sec.
Insp. vol. agent mixture
higher than 3 xMAC.
Insp. vol. agent higher than
alarm limit for 30 sec.
Inspiratory Sevoflurane con-
centration fell below lower
limit for 2 breath. cycles.
Volatile agent in inspiration
higher than setpoint.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Check Oxygen delivery set-
ting! Check fresh gas flow!
Check O
supply! Check
2
Oxygen delivery setting!
Check fresh gas flow!
Check SpO
concentration!
2
Check breathing pressure!
Call DrägerService.
Check insp. hose, tube and
filter.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Check DIVA settings and
fresh gas flow.
Check volatile agent deliv-
ery setting! Check fresh gas
flow!
Check volatile agent deliv-
ery setting! Check fresh gas
flow! Check function and
liquid level of DIVA. Check
CO
-sample line for leak-
2
age!
Use Fresh-gas control
mode.