Preventive maintenance

WARNING
Risk of faulty components
Device failures might result from wear and
material fatigue of components.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
The following table shows the service intervals:
User-specific intervals
AGS filter
IBF absorber filter
Filter of the endotracheal suction sys-
tem
Flow sensors
Plug of the bacterial filter in the en-
dotracheal suction system
Water trap (CO
Protect)
2
NIBP measurement
Instructions for use Zeus Infinity Empowered SW 2.n
Exchange if clogged.
Exchange after 20 operating hours.
Exchange every 14 days.
Exchange if calibration is no longer possible or in the event of er-
ror message Exp. flow sensor failure or Insp. flow sensor fail-
ure.
They may be incinerated with little pollution at temperatures over
800 °C.
Exchange after 2 years.
Replace in the event of contamination or the message Sample
line disconnected? (if the sample line is open and not kinked)
but at least every 4 weeks. Follow the hospital hygiene regula-
tions and relevant instructions for use for disposal.
to be checked annually by trained personnel