Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Caution
250
Power supply temp
high
Caution
000
Press.supp. time
limited
Note
000
Printer error
Caution
100
Protect sensor war-
mup
Note
255
Protect water trap
blocked
Caution
010
Protect water trap?
Caution
040
Pump 1 battery
empty
Note
040
Pump 1 battery low
Caution
200
Pump 1 failure
Note
120
Pump 1 incompati-
ble
298
Cause
Power supply temperature
too high.
Pressure support cycle was
aborted. The OFF criterion
(flow <25 % of max. insp.
flow) was not reached within
4 sec.
Could not access any printer.
Protection sensor is not
available. Agent delivery re-
quires a fresh gas flow
>3 L/min.
Protect water trap or gas
measurement system
blocked. Protect water trap
full or expired.
Protect water trap or gas
measurement system
blocked. Protect water trap
full or expired.
Infusion 1 stopped. Battery
of pump 1 discharged.
Battery nearly empty.
Infusion 1 stopped. Pump
detected internal failure.
Pump 1 hardware or soft-
ware version not compatible.
Remote control not possible.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Minimize power consump-
tion. Ventilate manually.
Switch off device as soon
as possible.
Call DrägerService.
Check for tube/mask leak-
age. Check patient settings.
Use Pressure Control.
Choose another printer, if
available.
Call DrägerService.
Latest after 1/2 hour the
protect sensor calibrates
again. After this the delivery
may be available again.
Use fresh gas flow
>3 L/min.
Check Protect water trap
and gas measurement sys-
tem – exchange, if neces-
sary. Exchange water trap
at least every 4 weeks.
Check Protect water trap
and gas measurement sys-
tem – exchange, if neces-
sary. Exchange water trap
at least every 4 weeks.
Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Reconnect IVDock to sys-
tem for power supply or ex-
change battery. Call
DrägerService.
Exchange pump 1. Call
DrägerService.
Control infusion at pump.
Replace pump for remote
control.