Dräger Zeus Infinity Empowered Instructions For Use Manual page 310

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Caution
000
SV tachycardia
Note
050
Sync fail
Caution
210
Sync. ventilation
failure
Warning
130
Syringe 1 empty
Caution
050
Syringe 1 nearly
empty
Caution
125
Syringe 1 unlocked
Warning
130
Syringe 2 empty
310
Cause
Superventricular tachycar-
dia: N or more consecutive
normal beats, with a beat-to-
beat rate greater than or
equal to the SVT setting.
Defibrillator sync. output
failed
Insp. and exp. flow sensor
failed. No triggering for as-
sisted ventilation available.
Infusion 1 stopped.
Syringe 1 empty.
Syringe 1 nearly empty.
5 min until end of infusion or
10 % volume left. (depend-
ing on infusion rate and type
of syringe)
Infusion 1 can not start while
clasp is unlocked.
Infusion 2 stopped.
Syringe 2 empty.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Check patient and take ap-
propriate action.
Call DrägerService.
Observe sensor error mes-
sages. Use alternative ven-
tilator mode.
Control syringe and patient.
Replace syringe, if neces-
sary. To drain syringe com-
pletely, confirm the green
button at the pump. (Only
possible, if 'Empty syringe
mode' is configured).
Control syringe. Prepare
filled syringe, if necessary.
Lock clasp at the pump 1.
Control right position of the
syringe. and start infusion
or bolus.
Control syringe and patient.
Replace syringe, if neces-
sary. To drain syringe com-
pletely, confirm the green
button at the pump. (Only
possible, if 'Empty syringe
mode' is configured).