Assembly and preparation

Before initial use

CAUTION
Before the device is used for the first time, all
parts which conduct breathing gas or which come
into direct contact with the patient must be treated
as described in the chapter "Care".
The device must have adjusted to current room
temperature sufficiently for it to be operating
within a temperature range of +15 °C to +35 °C.
Charging the battery for emergency op-
eration
Zeus IE has an integrated uninterruptible power
supply (UPS) which maintains the electrical supply
to the device in the event of mains failure for at least
30 minutes provided the battery is charged.
Switchover to battery operation takes place auto-
matically in the event of a failure. The Mains sup-
ply failure message appears.
The battery is automatically recharged if the mains
plug has been inserted.
A B
O
2
Air N
2
O
O
2
Air N
2
O
58
Charge the battery for 10 hours before the initial
use:
1 Insert the mains plug of the Zeus IE workstation
into the mains power socket.
The power supply must correspond to the voltage
indicated on the rating plate.
2 Check whether the circuit breaker on the right
next to the gas supply block is pressed.
The green LED (A) with the
Leave Zeus IE connected to the electrical power
supply for 10 hours. It does not need to be switched
on.
When the battery is charged, the green LED (B)
lights up with the

Installing flow sensors

WARNING
Risk of fire
Residual vapors of highly flammable disin-
fectants (e.g., alcohols) and deposits that
were not removed during reprocessing may
ignite when the flow sensor is in use.
– Ensure particle-free cleaning and disinfec-
tion.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor check for
visible damage and soiling, such as resid-
ual mucus, medication aerosols, and parti-
cles.
– Replace flow sensors when damaged,
soiled, or not particlefree.
Instructions for use Zeus Infinity Empowered SW 2.n
symbol lights up.
.