Setting mixed gas in the therapy bar on the
main screen
After closing the MixGas settings dialog window,
the most important therapy controls are displayed
in the therapy bar of the main page and can be set
there directly.
K
M
L
N
Setting the inspiratory O
1 Touch the (L) button.
2 Use the rotary knob to set and confirm the
value.
If 100% O
should be delivered:
2
3 Touch the 100% O
2
therap ycontrol turns yellow and the value 100
is displayed.
4 Confirm with rotary knob.
Setting the flushing flow:
1 Touch the Flush flow button (M).
2 Use the rotary knob to set and confirm the value
(max. 18 L/min).
The button color changes to dark green and dis-
plays the set value. The fresh-gas flow is increased
for 60 s to the set value. The text Flushing is dis-
played above the virtual flow tubes.
Interrupting the flush flow:
Touch the (N) button.
The flushing flow is interrupted and displays the set
MinFG Flow (closed or minimum fresh-gas flow).
Instructions for use Zeus Infinity Empowered SW 2.n
concentration:
2
button (K). The color of the
Operation
119