Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Warning
250
AW pressure high
Caution
250
AW pressure low
Warning
220
AW pressure nega-
tive
Note
175
AW pressure sen-
sor failure
Note
175
AW pressure sen-
sor failure
Note
170
Bag pressure sen-
sor fail
Note
150
Battery failure
Caution
020
Battery low
262
Cause
Protection system: airway
pressure higher than set-
point. Stenosis or patient
coughs.
Airway pressure too low.
Ventilation pressure not suf-
ficient. Leakage of low fresh
gas.
Mean airway pressure lower
than –2 cmH2O. Airway
pressure lower than
–10 cmH
O. Bronchus suc-
2
tioning. Inspiration reservoir
insufficient.
Airway pressure measure-
ment not plausible. Insuffi-
cient dynamic response.
Insp.+exp. pressure sensor
failed. No pressure monitor.
Ventilator controlled by tur-
bine speed.
Failure of manual vent. bag
pressure sensor. Automatic
Fresh-gas control not possi-
ble!
Battery of the uninterruptible
power supply failed.
The capacity of the uninter-
ruptible power supply is low
(<50 %).
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Observe pressure gauge.
Ventilate manually.
Check breathing circle leak-
age. Check for adequate
fresh gas flow.
Disconnect tube when suc-
tioning. Increase fresh gas
flow or use O
-flush.
2
Check airway pressure
sensors. Perform self test.
Terminate automatic
ventilation. Ventilate patient
manually. Call
DrägerService.
System automatically in-
creases fresh gas flow. Call
DrägerService.
Exchange battery. Call
DrägerService.
Restore power supply. Pre-
pare for man. ventilation.
Minimize power consump-
tion. Check battery capacity
display. If capacity is <10 %
the device may be switched
off!