Technical data

General information

Units of measurement for pressure
All specified tolerances apply for 20 °C (68 °F),
60 % relative humidity, and 1013 hPa
(760 mmHg).
The accuracies indicated below change according
to ambient pressure, temperature, and relative hu-
midity. If one of the ambient conditions is changed
up to the permissible limit, the accuracy of the cor-
responding value can change by up to 50 %. If
more than one of the ambient conditions are
changed, the accuracy may change by up to
100 %.
Example: Accuracy of a measured pressure value:
±4 % under standard conditions. At 10 °C, the ac-
curacy changes to ±6 % and at 10 °C and 20 %
relative humidity to ±8 %.
All patient-related volumes and flow values are
standardized to the conditions in the lung. (BTPS)
Percentages relate to the measured value, not to
the scale end.
If tolerances are stated both as a percentage and
as an absolute figure, it is always the higher figure
which applies!
396
1 hPa = 1 mbar = 1 cmH
100 kPa = 0.1 MPa = 1 bar = 1 kPa x 100
Instructions for use Zeus Infinity Empowered SW 2.n
O
2