Cleaning, disinfection, and sterilization

Care instructions for monitoring accessories

Patient cables

Only rub down patient cables with a cloth
soaked in soapy water.
Dry thoroughly with a lint-free cloth.
Patient cables can be disinfected with a cloth
soaked in diluted alcohol or an aldehyde-based
disinfectant.
Dry thoroughly with a lint-free cloth.

Reusable ECG electrodes

Remove any deposits from the electrodes regularly
with a small brush (toothbrush or the like). With re-
usable electrodes, the gel remnants must be re-
moved from the electrodes under running water
with a soft brush. Only rub down electrodes with a
cloth soaked in soapy water.
To disinfect, rub the electrodes with a cloth soaked
in diluted alcohol or an aldehyde-based disinfect-
ant. Dry thoroughly with a lint-free cloth.

NIBP cuff

Rub the cuff with a cloth dipped in soapy water, in
diluted bleach (1:10), an aldehyde-based disinfect-
ant or an alcohol or phenol solution.
CAUTION
The cuff may be dipped into a cleaning solution
but make sure the solution does not flow into the
hose opening or else the device will not operate
properly.
376

Hemo modules

WARNING
Danger of damaging the medical device
Do not immerse the device or its accessories
in liquids, or rinse them with liquids. If the de-
vice comes into contact with a liquid, discon-
nect it from the power supply immediately. Al-
low the device to dry completely for 24 to
48 hours. Contact service personnel to obtain
further information about device safety before
putting the device back into operation.
CAUTION
Do not autoclave hemo modules.

IBP pressure transducers

Pressure transducers and accessories must al-
ways be treated with extreme care. The pressure
transducer diaphragm must not be exposed to ex-
cessive pressure. Pressure transducers must not
be subjected to water, steam or hot-air sterilization
and must not come into contact with ether, chloro-
form or similar chemicals. The connection plug
must always be protected against moisture.
Follow the instructions supplied with the pressure
transducer for cleaning and sterilization!
Pressure transducer plates for the hemo
modules
The pressure transducer connecting plates of the
hemodynamic modules may be removed for clean-
ing and rinsed down in hot soap and water.
Instructions for use Zeus Infinity Empowered SW 2.n