Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Warning
255
Safety settings ac-
tive
Note
180
Sample line dis-
connected?
Warning
100
SC: Internal error
Warning
110
SC: Patient not
suitable
Caution
095
SC: Recovery
failed
Note
195
SC: Respiratory
drive might be suffi-
cient
Note
195
SC: Respiratory
drive sufficient
Warning
100
SC: Severe hypov-
entilation
Caution
085
SC: Spont. breath-
ing detected
Note
195
SC: Spontaneous
breathing insuffi-
cient
Caution
100
Service access al-
lowed
Caution
000
Setting not con-
firmed
306
Cause
Due to MIXER + DIVA fail-
ure, safety settings have
been activated.
Sample line leakage!
A software runtime error has
occured. SmartCare is not
available.
SmartCare cannot be contin-
ued. Patient values are not
within the etCO
and VT
2
ranges.
SmartCare could not induce
spontaneous breathing dur-
ing recovery.
The patient might be ready
for extubation.
The respiratory drive is suffi-
cient.
etCO
is too high.
2
Spontaneous breathing oc-
curred during controlled ven-
tilation.
Spontaneous breathing ac-
tivity of patient is insufficient.
Warning: The service access
via cable connection is al-
lowed! Ensure that no pa-
tient is connected.
Setting or change not con-
firmed within the timeout of
15 sec.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Check for additional mes-
sages. Check O
and DIVA
2
settings.
Check CO
sample line or
2
water trap -CO
-.
2
A software reset is required
before SmartCare can be
restarted. Press 'Software
reset?' button. The system
will reboot and be available
again after approximately
30 seconds.
Check patient data. Check
patient condition. Check
ventilation settings.
Check patient medication.
Check settings for Trigger.
Wait until patient is ready
for recovery.
Check the respiratory drive
of the patient to determine if
the patient can be extu-
bated.
Consider extubating the pa-
tient.
Check ventilation parame-
ters.
Check patient medication.
Allow spontaneous breath-
ing.
Check patient medication.
Check settings for Trigger.
After finishing service deac-
tivate 'Service access al-
lowance'!
Repeat last setting or
change if needed.