Maintenance
Replacement by DrägerService or a technician specifically trained for this purpose
Lead-gel battery UPS (2 pcs.)
Pressure controller (8417029) of the
gas supply block, (3 pcs.)
Sample gas return line filter
Filters upstream of pressure sensors Replace by trained personnel after 1 year.
Filter mat of the power unit
Diaphragms in the breathing system
(3 pcs.)
Sinter filter – gas supply block (3 pcs.) Replace after 2 years by DrägerService or a technician specifi-
Dust filter – device ventilation
Central supply pressure reducers –
fresh-gas mixer (3 pcs.)
Replacement by trained personnel
Filter mat for gas measuring module
NBP air filter in the Apollo drawer
O-rings (6870522) on water trap
holder, (2 pcs.)
O-ring (6870529) on water trap holder Replace after 2 years by trained personnel.
Small sealing rings (8602998) breath-
ing system, (6 pcs.)
Large sealing rings (M33729) breath-
ing system, (3 pcs.)
Guide sleeves (MK03310) DIVA me-
tering module, (2 pcs.)
Valve body (MK01580) DIVA meter-
ing module, (2 pcs.)
O-rings (MK02884) DIVA metering
module, (2 pcs.)
Distance washers (MK03284) DIVA
metering module, (2 pcs.)
Lithium batteries for the Diana and
Verona circuit boards
Lithium battery for the Dirona circuit
board (alternative)
388
After 3 years or the message Battery failure have it replaced by
trained personnel. Dispose of it in accordance with the local
waste disposal regulations.
Replace after 6 years by DrägerService or a technician specifi-
cally trained for this purpose.
Replace by trained personnel after 1 year.
Replace by trained personnel after 1 year.
Replace after 2 years by trained personnel.
cally trained for this purpose.
Annual visual inspection, to be replaced by trained personnel af-
ter 3 years.
Replace after 6 years by trained personnel.
Replace annually by trained personnel.
Replace after 2 years by trained personnel.
Replace after 2 years by trained personnel.
Replace after 3 years by trained personnel.
Replace after 3 years by trained personnel.
Replace after 3 years by trained personnel.
Replace after 3 years by trained personnel.
Replace after 3 years by trained personnel.
Replace after 3 years by trained personnel.
Replace after 6 years by trained personnel.