Classifi-
Pri-
Alarm text
cation
ority
Warning
045
VA delivery cutoff
Warning
045
VA delivery failure
Warning
045
VA delivery off
Warning
255
Vent+mixer+DIVA
cutoff
Warning
255
Vent. fibr. / Vent.
tach.
Note
100
Vent. turbine temp.
high
Warning
135
Ventilator failure
Caution
000
Ventr. bigeminy
Caution
000
Ventr. couplet
Warning
000
Ventr. tachycardia
Caution
000
Ventricular run
Instructions for use Zeus Infinity Empowered SW 2.n
Cause
DIVA failure. DIVA delivery
interrupted.
Failure of anesthetic agent
delivery system! Or delivery
terminated by protection sys-
tem. Delivery was switched
off!
Failure of anesthetic agent
delivery.
Failure of Ventilator, mixer
and DIVA. Mixer and DIVA
interrupted.
Ventricular fibrillation: The
monitor identifies a sinusoi-
dal waveform with fibrillation
characteristics.
Temperature of ventilator tur-
bine is too high.
Failure of automatic ventila-
tion: - Ventilator turbine
failed - Temperature of venti-
lator too high.
Sequence of beats with the
pattern: normal, PVC, nor-
mal, PVC, normal
Sequence of beats with the
pattern: normal, PVC, PVC,
normal
Ventricular tachycardia: N or
more PVCs are detected in
an interval T = (60 * (N – 1))
/ R, where N = VT count and
R = VT-rate
Ventricular run: Series of 3 to
N-1 consecutive PVCs with a
beat-to-beat rate > the VT
rate.
Fault – Cause – Remedy
Remedy
Continue anesthesia with
IV drug.
Call DrägerService.
Continue anesthesia with
IV drug.
Call DrägerService.
Continue anesthesia with
IV drug.
Call DrägerService.
Open O
Safety valve.
2
Monitor ventilation by
press. gauge.
Call DrägerService.
Check patient and take ap-
propriate action.
Reduce breathing pres-
sure! Ventilate manually!
Ventilate manually! Switch
to Fresh-gas control! Check
breathing system and tur-
bine. Call DrägerService.
Check patient and take ap-
propriate action.
Check patient and take ap-
propriate action.
Check patient and take ap-
propriate action.
Check patient and take ap-
propriate action.
315