Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Caution
125
Syringe 4 unlocked
Note
050
System restart re-
quired
Caution
050
System restart re-
quired
Warning
255
System restart re-
quired
Caution
150
T1 out of range
Caution
150
T2 out of range
Caution
000
Tachy
Note
050
Temp 1 hardware
fail
Caution
210
Temp 1 high
Caution
020
Temp 1 low
312
Cause
Infusion 4 can not start while
clasp is unlocked.
System has been running for
more than 7 days.
System has been running for
more than 14 days.
Internal system error.
Temperature value greater
or less than measuring
range
Temperature value greater
or less than measuring
range
Sinus tachycardia: N or more
consecutive normal beats,
with a beat-to-beat rate
>= TACH rate setting.
Temperature measurement
system failure.
Temperature exceeds upper
alarm limit.
Temperature exceeds lower
alarm limit.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Lock clasp at the pump 4.
Control right position of the
syringe. and start infusion
or bolus.
Shut down and restart the
device at the next opportu-
nity.
Shut down and restart the
device at the next opportu-
nity.
Shut down and restart the
device as soon as possible.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Check patient and take ap-
propriate action.
Call DrägerService.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.