Dräger Zeus Infinity Empowered Instructions For Use Manual page 297

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Classifi-
Pri-
Alarm text
cation
ority
Note
050
PA zeroing
Caution
100
Patient monitoring
fail
Caution
000
Pause
Caution
060
PEEP high
Caution
050
PEEP/P
failure
Note
050
Perform self test
Note
190
PGA fan failure
Caution
045
P
Caution
080
P
Caution
080
P
tained
Note
200
Power supply com-
munic. fail
Instructions for use Zeus Infinity Empowered SW 2.n
Cause
Mercury calibration in
progress.
Complete patient monitoring
component failed.
Sequence of two beats clas-
sified as normal or PVC, with
interval? pause rate value in
seconds (100 ms)
Measured PEEP 5 mbar
above setpoint for more than
10 breaths.
valve
PEEP/P
max
tive or the PEEP is higer than
the setting.
No self test performed within
24 hours.
Fan failure in patient gas
measuring module.
not attained
Due to a low fresh-gas flow
insp
or leakage the set inspiratory
pressure could not be at-
tained.
not attained
Due to time settings the set
insp
inspiratory pressure could
not be attained.
lim/VT not at-
Due to time settings or P
max
limitation the set volume
could not be attained. Pa-
tient compliance or resist-
ance have changed.
Communication failure of
power supply. Battery, temp.
and voltage monitor not
available.
valve is defec-
max
max
Fault – Cause – Remedy
Remedy
Complete calibration before
you begin monitoring the
patient.
Press reset button, if sys-
tem repair is desired. Sys-
tem will not be available for
appr. 30 sec.!
Check patient and take ap-
propriate action.
Check breathing system.
Exchange autom exp.
valve. Call DrägerService.
Check breathing pressure
and volume. Ventilate man-
ually. Check breathing sys-
tem for dust particles. Call
DrägerService.
Perform self test.
Call DrägerService.
Increase fresh-gas flow,
seal leaks. Provide suffi-
cient ventilation.
If necessary, adjust ventila-
tion settings for P
, Pres-
insp
sure Support/PEEP, slope,
frequency and I/E. Provide
sufficient ventilation.
If necessary, adjust ventila-
tion settings for slope, fre-
quency and I/E. Check
P
settings. Provide suffi-
max
cient ventilation.
Use system only in case of
emergency.
Call DrägerService.
297