Configuration
Printer
The General settings > Printer page opens.
Basic settings
The list of available printers (defined in the operat-
ing system) appears. The settings are made by
DrägerService.
The printer selected here only prints the current
screen.
The network printer, which is connected via the In-
finity network, can print out all graphic and tabular
trends.
CAUTION
Danger resulting from incorrect connection of the
printer
If incorrect printer cables are used, these cables
may cause the leakage current to exceed the per-
missible requirement for medical devices and
may present a hazard if the patient comes into
contact with them.
If the printer is connected to the LAN port, use
printer cable MK04638. Connect the printer cable
corresponding to the labeling for the cable on the
printer.
340
CAUTION
If office equipment or components such as a
printer or network switch are used in conjunction
with Zeus IE, the products must conform to pro-
tection class 1 and meet the requirements of
IEC 60950-1.
Such devices or components must not be con-
nected to the auxiliary Zeus IE sockets. They may
only be connected via Ethernet.
These devices or components must be installed
outside the patient environment (1.5 m round the
patient bed or operating table) and must not be
touched by the user at the same time as the pa-
tient.
If ignored, it can put the patient and user at risk.
With regard to the following printers, Dräger has
given approval confirming correct data exchange:
– Zeus IE with HP Laser Jet 4250N
– Zeus IE with HP Laser Jet 4015N
– Zeus IE with HP Laser Jet 600
CAUTION
Do not touch electrical contacts of connected data
cables and patients simultaneously. The electric
data cables could place the patient at risk or
measurement signals could be affected.
CAUTION
If additional medical devices are plugged into the
auxiliary sockets, the user must ensure that the
requirements of IEC 60601-1-1 are met.
Plugging medical devices into the auxiliary sock-
ets can bring about a considerable increase in dis-
charge current above permitted levels.
If ignored, it can put the patient and user at risk.