Technical data
Pulse oximetry (SpO
SpO
accuracy
2
424
) (Cont.)
2
Measuring accuracy in Adult mode
SpO
2:
0 to 69 % not specified
70 to 100 % sensor-specific, specified as follows:
(2, 3)
Nellcor
:
OxiMAX MAX-A, OxiMAX MAX-AL, Oxi-
MAX MAX-P, OxiMAX MAX-N, OxiMAX
MAX-I
Nellcor:
DS100A
(2, 3)
Masimo
:
LNOPADT, LNOPPED, LNOPNEO,
LNOPNEO SS, LNOP-YI
Masimo:
LNOP-DCI, LNOP-DCIP, NR125
EAR
Pulse:
±3 bpm or ±3 % (whichever is higher)
Measuring accuracy in Neonatal mode
SpO
2:
0 to 69 % not specified
70 to 100 % sensor-specific, specified as follows:
Nellcor:
N-25, OxiMAX MAX-N
Masimo:
LNOPNEO, LNOPNEO SS, LNOP-YI
Pulse:
±3 bpm or ±3 % (whichever is higher)
Instructions for use Zeus Infinity Empowered SW 2.n
(1)
:
±2
±3
±2
±2
±3.5
(1, 2, 3, 4)
:
±3
±3