Dräger Zeus Infinity Empowered Instructions For Use Manual page 273

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Classifi-
Pri-
Alarm text
cation
ority
Warning
255
Fresh gas in-
tern/extern?
Warning
044
Fresh-gas flow
< 3 L/min
Note
005
Gas supply sens.
failure
Note
050
GP1 check zero
Caution
090
GP1 D high
Caution
090
GP1 D low
Instructions for use Zeus Infinity Empowered SW 2.n
Cause
Failed to switch between ex-
ternal and internal fresh gas
outlet. No fresh gas is deliv-
ered to the requested
breathing system!!!
DIVA protection system
failed and fresh-gas flow
lower than 3 L/min.
Sensor for cylinder pressure
failed.
Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Invasive pressure value out-
side alarm limits.
Invasive pressure value out-
side alarm limits.
Fault – Cause – Remedy
Remedy
Check fresh gas flow and
anesthetic delivery. Switch
again between internal and
external outlet. If fresh gas
flow is not available use
emergency breathing bag.
Call DrägerService.
Set fresh gas flow higher
than 3L/min. NOTE:At
lower flows DIVA will be ter-
minated.
Check sensor plug. Ex-
change high pressure sen-
sor. Call DrägerService.
Zero the transducer.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
273