Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Caution
090
NIBP D high
Caution
090
NIBP D low
Note
050
NIBP hardware
fault
Note
050
NIBP hose un-
plugged
Caution
200
NIBP inflation fail
Caution
090
NIBP M high
Caution
100
NIBP M low
Caution
200
NIBP open line
292
Cause
Non invasive pressure value
outside alarm limits.
Non invasive pressure value
outside alarm limits.
Non invasive blood pressure
measurement failed.
NIBP hose is unplugged.
Systolic pressure is higher
than inflation limit. Pump or
valves have been energized
for longer than 2 minutes
(90 sec. neonatal mode).
Non invasive pressure value
outside alarm limits.
Non invasive pressure value
outside alarm limits.
Inflation time during cuff in-
flation cycle is too long or in-
flation rate is too low.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Call DrägerService
Reconnect the hose.
Select higher inflation limit
setting. Check hose and
cuff for obstructions. Dis-
connect and reconnect cuff.
Replace if necessary. Call
DrägerService.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check to ensure that the
hose and cuff are properly
connected to the monitor.
Check the cuff size to be
suitable for the patient.