Invasive Blood Pressure Measurement Ibp - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Invasive blood pressure measurement
IBP
Signals for invasive blood pressure monitoring
(IBP) are received via a Y-cable, one or more he-
modynamic modules or via a combination of these
devices.
A pressure transducer converts pressures into
electronic signals for the Zeus IE.
These pressure signals are filtered to eliminate ar-
tefacts.
Zeus IE detects individual beats by calculating
threshold values from the running average of the
systolic and diastolic pressures.
IBP parameters
Par.
Pressure
ART
Arterial
LV
Left ventricular
PA
Pulmonary arterial
RV
Right ventricular
CVP
Central venous
RA
Right atrial
LA
Left atrial
ICP
Intracranial
GP1
Unspecific, 1
GP2
Unspecific, 2
Instructions for use Zeus Infinity Empowered SW 2.n
Measured value
Systolic, diastolic, mean
Systolic, diastolic, mean
Systolic, diastolic, mean
Systolic, diastolic, mean
Medium
Medium
Medium
Medium
Systolic, diastolic, mean
Systolic, diastolic, mean
Monitoring
Measuring range
–50 to 400 [mmHg]
–50 to 400 [mmHg]
–50 to 400 [mmHg]
225

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents