Cleaning, disinfection, and sterilization
WARNING
Risk of fire
Residual vapors of highly flammable disin-
fectants (e.g., alcohols) and deposits that
were not removed during reprocessing may
ignite when the flow sensor is in use.
– Ensure particle-free cleaning and disinfec-
tion.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor check for
visible damage and soiling, such as resid-
ual mucus, medication aerosols, and parti-
cles.
– Replace flow sensors when damaged,
soiled, or not particlefree.
CAUTION
Risk of failure of flow measurement
Improper reprocessing and soiling, such as de-
posits or particles, may damage the flow sensor:
– No machine cleaning or disinfection of the
sensor insert
– No plasma sterilization or radiation steriliza-
tion
– No compressed air
– No water jets, compressed air, brushes, or
similar to be used on the sensor insert
– No ultrasonic bath
– Clean and disinfect the flow sensor in accord-
ance with the corresponding instructions for
use.
– For disinfecting the flow sensor use only clean
disinfectant solutions.
370

Sterilization

Sterilization eliminates living microorganisms from
semi-critical medical devices and dries residual wa-
ter in the interior of components.
Sterilize only components that have been
cleaned and disinfected.
For sterilization, use a vacuum steam sterilizer (in
accordance with DIN EN 285), preferably with frac-
tional vacuum.
CAUTION
These medical devices have been designed,
tested, and manufactured exclusively for single
use. Disposable products must not be reused, re-
processed, or sterilized. Reuse, reprocessing, or
sterilization may lead to failure of the medical de-
vices and injury to the patient.
Sterile accessories
CAUTION
Do not used sterile packed accessories if the
packing has been opened or damaged or there
are other indications of non-sterility. Reprocess-
ing and re-sterilization of disposable articles are
not permitted.
WARNING
Do not allow any alcohol or alcoholic agents
to get into the sample line openings! Alcohol
may distort the measuring results.