1 Turn the CO
absorber (A) counterclockwise
2
and remove downwards.
2 Remove the IBF absorber filter (B) using the re-
cessed handles in the filter.
3 Empty out the used soda lime and dispose of
according to the soda lime instructions for use.
4 Fill the CO
absorber with fresh soda lime up to
2
the upper mark.
5 Insert new IBF absorber filter. Only use undam-
aged filters, as exterior damage to the filter de-
creases protection!
6 Attach CO
absorber to the breathing system
2
from the bottom and turn clockwise until it en-
gages.
Instructions for use Zeus Infinity Empowered SW 2.n
A
Using disposable CO
Optional CLIC adapter
The CLIC adapter makes it possible to use
Drägersorb CLIC 800 Plus or
Drägersorb CLIC Free disposable CO
see instructions for use of disposable CO
ers.
No IBF filter is required if using the recommended
cartridges (dust filter is already integrated in the
cartridge).
Fitting the CLIC adapter
Insert the CLIC adapter into the breathing sys-
tem from below and turn clockwise as far as it
will go. Make sure that the unlock button (A)
is facing the front and visible to the
user.
Assembly and preparation