Classifi-
Pri-
Alarm text
cation
ority
Note
050
Temp 1 unplugged
Note
050
Temp 2 hardware
fail
Caution
200
Temp 2 high
Caution
020
Temp 2 low
Note
050
Temp 2 unplugged
Caution
080
Tidal vol. high (P
ltd.)
Caution
080
Tidal volume high
Note
100
Tidal volume high
Caution
045
Tidal volume not at-
tained
Note
210
Time synchroniza-
tion
Instructions for use Zeus Infinity Empowered SW 2.n
Cause
Cable defective or un-
plugged.
Temperature measurement
system failure.
Temperature exceeds upper
alarm limit.
Temperature exceeds lower
alarm limit.
Cable defective or un-
plugged.
At minimum airway pressure
min
tidal volume is higher then
set. – Breathing settings not
adequate. – Patient breathes
spontaneously.
Pressure Control: Upper limit
of insp. tidal volume has
been exceeded multiple
times. Volume Control:
VTset x 1.3 exceeded.
Pressure Control: Upper limit
insp. tidal volume reached.
Volume Control:
VTset x 1.3 exceeded.
Due to fresh-gas deficiency
the set volume could not be
attained.
The clock has been synchro-
nized with system time.
Fault – Cause – Remedy
Remedy
Check equipment and re-
place if necessary.
Call DrägerService.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Check equipment and re-
place if necessary.
Change ventilation mode or
VT settings.
Observe minute volume.
Check alarm limit of insp.
tidal volume and P
insp
ting.
Check alarm limit of
insp.tidal volume. Switch off
AutoFlow mode.
Increase fresh-gas flow,
seal leaks. Provide suffi-
cient ventilation.
This message will disap-
pear automatically after
15 seconds.