Water Trap For Gas Sensors - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Water trap for gas sensors

Zeus IE is equipped with two water traps to protect
the device monitoring system:
– Protect: Water trap in the internal sample line
– CO
: Water trap for the patient sample line
2
CAUTION
Water traps must be changed after 4 weeks of use
at the latest. If they are used for a longer period,
the diaphragm could become permeable, allowing
water and bacteria into the measuring system.
Contamination has a negative effect on gas anal-
ysis, which can fail completely.
CAUTION
Avoid using aerosols in the breathing system. Do
not use a water trap in conjunction with drug neb-
ulization. Do not allow any alcohol or deter-
gent/disinfectant to come into contact with the in-
side of the water trap. Do not wash, rinse or
sterilize a water trap. Such substances can dam-
age the diaphragm and the measuring system.
CAUTION
The presence of organic detergents or gases con-
taining freon can affect the measuring accuracy of
the gas measurement.
Instructions for use Zeus Infinity Empowered SW 2.n
Cleaning, disinfection, and sterilization
Emptying and replacing water traps
Note the instructions for use of the water trap.
Empty the water trap:
– When the filling level has reached the upper
mark.
– When a message appears on the screen.
Replace the water trap:
– When the maximum period of use is reached.
Emptying the water trap
1 Remove hose (A) and pull off water trap (B).
D
2 Insert an empty syringe (C) without canula, at
least 20 mL, into the blue connector (D).
3 Extract water, remove the syringe and dispose
of the full syringe as infectious hospital waste.
Observe the hospital's hygiene regulations!
4 Push the water trap (B) into the holder again un-
til it engages and connect hose (A).
A
B
C