Checking The Apl Valve; Emergency Ventilation Bag; Checking The Anesthetic Gas Scavenging; System Agss - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Startup

Checking the APL valve

B
(for manual checking in cases of emergency)
1 Adjust the pressure to 20 mbar (A).
2 Manually close the filter on the Y-piece or con-
nect the test lung.
3 Manually ventilate with breathing bag.
4 Check the pressure limits on the mechanical
pressure gauge (B).

Emergency ventilation bag

Example: Dräger Resutator 2000 (not shown).
The emergency ventilation bag is present on
the device and its functionality has been
checked.
92
A

Checking the anesthetic gas scavenging

system AGSS

The anesthetic gas scavenging system (AGSS)
must conform to ISO 8835-3.
WARNING
Risk of patient injury
A vacuum can be created in the breathing sys-
tem and in the patient's lung if the side open-
ings on the receiving system are covered.
Always make sure that the side openings on
the receiving system are not blocked.
A
1 Transfer hose (A) is connected to the nozzle on
the back of the device and to the receiving sys-
tem (AGS).
2 The scavenging hose is connected to the re-
ceiving system, the plug (B) is in the Dräger wall
socket, which has a green indicator.
3 The float in the AGS hangs between the two
markings.
Follow the instructions for use of the anesthetic
gas receiving systems (AGS) and the anes-
thetic gas scavenging system (AGSS).
Instructions for use Zeus Infinity Empowered SW 2.n