Monitoring
Connection for IBP accessories
For invasive blood pressure measurement, four al-
ternatives are available on Zeus IE:
– Y-cable
– Hemo2 Pod
– Hemo4 Pod
– HemoMed Pod
– Quad Hemo Pod
Y-cable
The Y-cable can be used to connect two IBP pres-
sure transducers to measure two invasive blood
pressures. The Y-cable is connected to the
HEMOMED 1 connection on the Zeus IE connec-
tion field, see page 197.
C
A Connectors for the connection to Zeus IE
B Y-cable
C Intermediate cable
D IBP pressure transducer
Hemo2 Pod, Hemo4 Pod, HemoMed Pod and
Quad Hemo Pod
E
F
202
B
A
D
optionally without
SmartPod SpO
2
E
F
Hemodynamic monitoring modules which perform
the following measurements with only one monitor
connection:
– Cardiac output
– 2 invasive blood pressure measurements
(Hemo2) or
– 4 invasive blood pressure measurements
(Hemo4, HemoMed and Quad Hemo)
Hemo2 and Hemo4 must be connected to an
Aux/Hemo 2/3 (F) on Zeus IE.
The HemoMed Pod must be connected to the
HEMOMED 1 connection (E) on Zeus IE.
In contrast to the Hemo2 and Hemo4 Pods, the
HemoMed Pod has no LCD displays on the module
and no temperature connections.
The Quad Hemo Pod has two temperature connec-
tions and no LCD display.
Hemo2 Pod, Hemo4 Pod and HemoMed Pod
The modules have three buttons:
G The SmartZero
of all pressure transducers connected to the
module which are open to atmospheric pres-
sure.
H The C.O. Start key starts measuring cardiac
output.
I
The Wedge key starts measuring pulmonary
wedge pressure.
Instructions for use Zeus Infinity Empowered SW 2.n
G H
I
key performs the zeroing