Startup
Terminating the self test:
Touch the Cancel button (H).
If the self test is canceled during the test sequence,
the canceled message appears instead of com-
pleted.
If the self test failed, the Test details dialog window
remains open.
There is an audible tone at the end of the self test.
98
I
The system standby screen displays the following
information (I):
– Exchange water traps at least every
4 weeks.
– Last self test date and time
– Run self test at least every 24 hours
– Last test result (with colored background).
Test of the alarm system
The alarm system is automatically checked during
the self test. Additionally, the loudspeaker is auto-
matically checked when a new case is started.
Instructions for use Zeus Infinity Empowered SW 2.n