Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Note
020
LL electrode fail
Note
050
LV check zero
Caution
090
LV D high
Caution
090
LV D low
Note
025
LV hardware fail
Caution
090
LV M high
Caution
100
LV M low
288
Cause
Lead-off condition for indi-
cated lead detected: -Broken
cable – Loose lead wire
-Faulty lead wire – Wrong
lead -Dried out gel -Bad con-
tact.
Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Invasive pressure value out-
side alarm limits.
Invasive pressure value out-
side alarm limits.
Invasive blood pressure
hardware failure.
Invasive pressure value out-
side alarm limits.
Invasive pressure value out-
side alarm limits.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Zero the transducer.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check hardware and re-
place. Call DrägerService.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.