When Filling And Emptying - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Fault – Cause – Remedy
Fault
Smell of anesthetic agent dur-
ing or after transport/storage.

When filling and emptying

Fault
Anesthetic agent flows out of
the drain port.
Anesthetic agent is escaping
at the cylinder thread.
Anesthetic agent is escaping
at the filling device.
Anesthetic agent flows out of
the overflow hole.
The locking bolt cannot be
pulled out.
Anesthetic agent does not
flow out when emptying.
318
Cause
In the event of an extreme rise in
temperature and/or drop in at-
mospheric pressure, anesthetic
agent vapor can escape or liquid
anesthetic agent can leak as a re-
sult of pressure in the metering
module.
Cause
Drain valve not closed.
The filling adapter is not screwed
firmly onto the bottle.
The seal in the screw cap of the
filling adapter is missing or dam-
aged.
The filling adapter is not com-
pletely inserted or the lever is not
firmly tightened.
The lever is not pressing firmly
enough onto the filling adapter.
The filling adapter is damaged.
The seal of the filling device is
damaged.
The metering module is filled
above maximum.
The lever is not sufficiently
opened or the lever is incorrectly
adjusted.
The locking bolt/cap is closed.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Do not inhale the anesthetic
agent vapor. Air the room. Allow
the metering module to settle at
room temperature. Observe the
application range of the filled
metering module.
Remedy
Close drain valve.
Screw the filling adapter on
firmly.
Check the seal.
Contact DrägerService
Loosen the lever, insert the filling
adapter completely, tighten the
lever.
Have the lever readjusted.
Contact DrägerService
Use another filling adapter.
Contact DrägerService
Check the seal of the metering
module.
Contact DrägerService
Empty metering module up to
the maximum mark.
Open the lever further or have it
readjusted.
Open the locking bolt/cap.
Close again tightly after empty-
ing.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents