Installing Vaporizers; Installing The Manual Breathing Bag Arm; Connecting The Breathing Hoses - Dräger Apollo Operating Instructions Manual

Inhalation anesthesia machine software 3.2n
Hide thumbs Also See for Apollo:
Table of Contents

Advertisement

Chapter 4 - System Setup
5. Make all necessary connections between the
suction bottle, suction regulator, and piped
vacuum system as specified in the Instructions
for Use provided with the bottle assembly.

Installing Vaporizers

Install vaporizers as directed in the appropriate
Instructions for Use supplied with the vaporizers
available for use with the Apollo.
Installing the Manual Breathing
Bag Arm
WARNING !
Breathing bags used on the Apollo must comply
with current ANSI standards.
1. Slide the bag arm assembly onto the breathing
bag port on the side of the breathing system (1 in
Figure
32).
2. Tighten the two thumb screws (2 in
secure.
3. Attach the 90° fitting to the end of the bag arm (3
in
Figure
32), and attach the breathing bag to the
other end of the fitting.

Connecting the Breathing Hoses

WARNING !
Breathing hoses used on the Apollo must
comply with current ANSI standards.
WARNING !
Do not use antistatic or conductive breathing
hoses. There is a risk of burns when using high
frequency electro-surgical equipment.
50
Figure 32. Breathing Bag Arm Connection
Figure
32) to

Installing Vaporizers

3
2
2
1
Apollo Operator's Manual