Connecting the Breathing Hoses

NOTE:
Take care not to damage the breathing hoses.
When connecting and disconnecting, always hold
the breathing hoses by the end sleeve, not by the
spiral reinforcement. Otherwise, the spiral rein-
forcement may be torn loose.
Breathing hoses with a damaged spiral reinforce-
ment can kink or become occluded.
Before each use, check the breathing hoses for
damage.
WARNING
Risk of use of toxic or incompatible materials
Breathing hoses used on the Fabius OS must
comply with current ANSI standards.
WARNING
Do not use antistatic or conductive breathing
hoses. There is a risk of burns when using
high frequency electrosurgical equipment.
Fabius OS has no components containing
latex.
For latex-free use:
Use latex-free breathing bag and breathing
hoses!
1 Push patient breathing hoses onto both the
inspiratory and expiratory connectors or onto
optional microbial filters.
2 Connect both patient breathing hoses to the
Y-piece.
3 Push the bag onto the end of the breathing bag
extension.
WARNING
Risk of strangulation
Use caution when connecting the patient.
Instructions for Use Fabius OS SW 3.n
3
2
1
Assembly
55