Dismantling The Breathing System; Removing The Flow Sensor; Removing The Absorber And Bellows - Dräger Julian Instructions For Use Manual

Anaesthetic workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Dismantling the breathing system

7 Fold up the writing top.
8 Pull the catch and at the same time pull out the slot-in
module.
9 Hold breathing system by the handle and pull
upwards to remove.

Removing the flow sensor

1 Unscrew the expiration port.
2 Remove the flow sensor.
The flow sensor cannot be disinfected / cleaned in a
washing machine, nor can it be sterilized in high-
temperature steam.
The flow sensor must be disinfected in 70 % ethanol
solution for approx. 1 hour.
Leave the sensor to dry in air for at least 30 minutes,
otherwise the remaining alcohol can damage the
sensor when calibrated.
The flow sensor can be reused as long as it can be
calibrated successfully.

Removing the absorber and bellows

3 Turn the absorber counterclockwise and pull it
downwards.
Empty the soda lime; it can be disposed of with
ordinary domestic waste.
4 Remove the insert from the absorber.
5 Turn the container of the bellows counterclockwise
and pull it downwards.
6 Remove the bellows.
Prepare the absorber, container of the bellows and
the bellows for disinfection and cleaning in a washing
machine.
7
9
2
4
Care
Dismantling components
8
1
MAX
6
3
5
99