Assembly

Activating the Battery

The Fabius plus anesthesia workstation is shipped
with the battery fuse disconnected in order to pre-
vent discharge during shipment and storage prior to
installation.
Remove the battery fuse from the top drawer of
the machine.
Remove the battery fuse from its packaging.
1 Insert the battery fuse into the battery fuse
holder. Turn the fuse holder 1/4-turn clockwise
until it is snug.
Fitting the CO
Absorber on the
2
Compact Breathing System
Fill the absorber with fresh CO
Dräger recommends the use of
Drägersorb 800 Plus or Drägersorb FREE. For
detailed information on filling and installing the
reusable absorber, see page 185.
CAUTION
Risk of unintended perioperative awareness
Make sure that during ventilation the absorber
canister is tightly connected to the COSY.
NOTE
Ensure that no CO
absorbent dust/particles have
2
been deposited between the gaskets and sealing
surfaces. Such dust and particles can cause leaks
in the system.
2 Fit the absorber canister into position below the
breathing system and turn it counterclockwise
as far as possible.
44
absorbent.
2
1
2
Instructions for Use Fabius plus SW 3.n