Fitting The Co Absorber On The Compact Breathing System; Breathing System - Dräger Fabius Instructions For Use Manual

Mri
Table of Contents

Advertisement

Fitting the CO
Absorber on the
2
Compact Breathing System
Fill the absorber with fresh CO
Dräger recommends the use of
Drägersorb 800 Plus or Drägersorb FREE. For
detailed information on filling and installing the
reusable absorber, see page 189.
CAUTION
Risk of unintended perioperative awareness
Make sure that during ventilation the absorber
canister is tightly connected to the COSY.
NOTE
Ensure that no CO
absorbent dust/particles have
2
been deposited between the gaskets and sealing
surfaces. Such dust and particles can cause leaks
in the system.
1 Fit the absorber canister into position below the
breathing system and turn it counterclockwise
as far as possible.
Instructions for Use Fabius MRI SW 3.n
absorbent.
2
Assembly
1
51