Connecting or exchanging consumables

WARNING
Risk of high inspiratory CO
Use of soda lime over prolonged periods can
increase the inspiratory CO
Check the color of the soda lime regularly
and exchange as needed, especially if the
inspiratory CO
value increases
2
unexpectedly.
Single-use CO
absorber
2
 Connect or replace the CLIC absorber or the
Infinity ID CLIC absorber.
Strictly observe the instructions for use of the
single-use CO
absorber.
2
WARNING
Risk of insufficient ventilation
If the CO
absorber is not correctly locked
2
into place, system leakage may occur.
After mounting and replacing, make sure the
CO
absorber is firmly locked into place.
2
Reusable CO
absorber
2
As an alternative to single-use CO
reusable CO
absorber may also be used.
2
CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for
eyes, skin, and airway.
Handle this absorption material carefully and do
not spill it.
Instructions for use Perseus A500 SW 1.1n
values
2
values.
2
absorbers, the
2
Dismounting and emptying
B
1 Turn the CO
absorber (A) clockwise and
2
remove it from below.
2 Remove and dispose of the soda lime dust
filter (B).
3 Empty used soda lime and dispose of
according to the instructions for use.
C
4 If it is necessary to clean the absorber insert
(C), remove the absorber insert from the
absorber container. Leave the inner and outer
sealing rings on the absorber insert.
Assembly and preparation
A
65